Clinical Analytical Instruments Original Mindray CBC First Hematology Analyzer With Word Alarm Information BC-30S
Principles: Impedance method for WBC, RBC and PLT counting Cyanide free reagent for hemoglobin test.
Performance
Parameter |
Linearity Range |
Precision (CV %) |
Carryover |
WBC(109/L) |
0-200 |
<=3.5% (4.0-6.9) |
<=0.5% |
|
|
<=2.0% (7.0 -15.0) |
|
RBC(1012/L) |
0-8.00 |
<=1.5% (3.5-6.5) |
<=0.5% |
HGB(g/L)
|
0-280
|
<=1.5% (100 -180)
|
<=0.5%
|
MCV(fL)
|
|
<=1.0% (70 -110)
|
|
PLT(109/L) |
0-4000 |
<=5.0% (100 -149) |
<=1.0% |
|
|
<=4.0% (150 -500) |
|
Parameters21 parameters: WBC, Lymph#, Mid#, Gran#, Lymph%, Mid%, Gran%, RBC, HGB, HCT, MCV, MCH, MCHC, RDW-CV, RDW-SD, PLT, MPV, PDW, PCT, P-LCR, P-LCC 3 histograms for WBC, RBC and PLT.
Reagent
- M-30D DILUENT.
- M-30CFL LYSE.
- PROBE CLEANSER.
Sample
- Volume Prediluted mode 20uL.
- Whole blood mode 9u L.
Throughput : 70 samples per hour.
Display : 10.4 inch TFT Touch Screen.
Data Storage Capacity: Up to 500,000 results including numeric and graphical information.
Communication
- LAN Port supports HL7 protocol.
- Support bi-directional LIS.
Interface : 4 USB port (for external printer, software upgrade, barcode reader, WIFI adapter,keyboard and mouse), LAN port (1).
Printout : Thermal recorder, 50 mm wide paper, various printouts formats External printer optional.
Operating Environment
- Temperature: 10°C~40°C.
- Humidity: 10%~90%.
- Air pressure: 70kPa-106kPa.
Power Requirement
- 100V-240V.
- <=300VA.
- 50Hz/60Hz.
Dimension and Weight
- Dimension: Depth(410 mm) x width(300 mm) x height(400 mm).
- Weight: <=20Kg.